Correlation Using an automated analyzer and a reference method based on the procedure of VandenBerg and Muller resulted in the following regression statistics: |
|
Method Performance Characteristics Sensitivity: The sensitivity of this method is 0.040 - 0.051 absorbance units per mg/dL. Linear Range: In this method there is no significant nonlinearity over the range of 0-30mg/dL. Precision: Within-run and day-to-day precision is summarized below. |
|
Intended Use For IN VITRO diagnostic use in the quantitative determination of Total Bilirubin in serum or plasma using manual or automated applications.
Method Principle Bilirubin from the serum sample mixes with diazotized sulfanilic acid to form an azo-Bilirubin complex. The increase in absorbance is monitored at 550nm. The reaction scheme illustrates the reaction that occurs in this method. |
|
TO CONTACT US:
CATACHEM, INC. 955 Connecticut Avenue Suite No. 4106 Bridgeport, CT 06607
Phone: 203-335-9277 Fax: 203-335-9977
Email: catachem@catacheminc.com
URL: www.catacheminc.com |


|
T. Bili |
Within-Run Precision |
Total Precision |
||
|
MEAN |
SD |
CV |
SD |
CV |
|
mg/dL |
mg/dL |
% |
mg/dL |
% |
|
0.45 |
0.09 |
* |
0.12 |
* |
|
13.15 |
0.11 |
0.86 |
0.16 |
1.20 |
|
26.50 |
0.16 |
0.61 |
0.39 |
1.50 |
|
Correlation Data |
|
|
Parameter |
Data Observed |
|
N |
108 |
|
Range |
0.30—16.7 |
|
Regression |
Y = 0.982x + 0.13 |
|
Correlation |
r = 0.999 |
|
Sy.x |
0.22 |