Correlation A comparison of this method using an automated analyzer and a reference method based upon the hexokinase-G6PDH reaction resulted in the following regression statistics. |
|
Method Performance Characteristics Sensitivity: 0.0008 - 0.0012 absorbance units per mg/dL. Linear Range: 0-600 mg/dL. Precision: Within-run and day-to-day precision is summarized below. |
|
Intended Use For IN VITRO diagnostic use in the quantitative determination of γ-Glucose in serum using manual or automated applications.
Method Principle The enzyme hexokinase catalyzes the phosphorilation of glucose in the presence of ATP and magnesium ions. The resultant glucoase-6-phosphate is then oxidized to 6-phosphoglucono lactone with concomitant reduction of nicotinamide adenine dinucleotide (NAD). The amount of NADH produced is proportional to the glucose present in the serum sample and it is quantitated at 340nm. The reaction scheme illustrates the reactions that occur in this method. |
|
TO CONTACT US:
CATACHEM, INC. 955 Connecticut Avenue Suite No. 4106 Bridgeport, CT 06607
Phone: 203-335-9277 Fax: 203-335-9977
Email: catachem@catacheminc.com
URL: www.catacheminc.com |
|
Glucose |
Within-Run Precision |
Total Precision |
||
|
MEAN |
SD |
CV |
SD |
CV |
|
mg/dL |
mg/dL |
% |
mg/dL |
% |
|
58 |
1.30 |
2.40 |
3.10 |
5.20 |
|
289 |
1.80 |
0.70 |
7.40 |
2.50 |
|
475 |
3.50 |
0.70 |
5.00 |
1.00 |
|
Correlation Data |
|
|
Parameter |
Data Observed |
|
N |
160 |
|
Range (mg/dL) |
60-472 |
|
Regression |
Y = 0.993 + 2.0 |
|
Correlation |
r = 0.999 |
|
Sy.x |
2.9 |