Correlation A comparison of this method using an automated analyzer and a reference method based upon the molybdate reaction resulted in the following regression statistics: |
|
Preparation Of Working Reagent Catachem Inorganic Phosphorus Reagent is packaged in ready-to-use form. No preparation is required.
Reagent Storage And Stability Catachem Inorganic Phosphorus reagent is stored at room temperature (18-26°C). When stored as directed, the reagent is stable until the expiration date stated on the label.
Method Performance Characteristics Sensitivity: 0.040-0.051 absorbance units per mg/dL. Linear Range: 0-10mg/dL. Precision: Within-run and day-to-day precision is summarized below. |
|
Intended Use For IN VITRO diagnostic use in the quantitative determination of Inorganic Phophorus in serum using manual or automated applications.
Method Principle Phosphorus from the serum sample mixes with the molybdate to form the phosphomolybdate complex. The increase in absorbance is monitored at 340nm. The reaction scheme illustrates the reaction that occurs in this method. |
|
TO CONTACT US:
CATACHEM, INC. 955 Connecticut Avenue Suite No. 4106 Bridgeport, CT 06607
Phone: 203-335-9277 Fax: 203-335-9977
Email: catachem@catacheminc.com
URL: www.catacheminc.com |
|
IP |
Within-Run Precision |
Total Precision |
||
|
MEAN |
SD |
CV |
SD |
CV |
|
mg/dL |
mg/dL |
% |
mg/dL |
% |
|
1.70 |
0.11 |
6.20 |
0.14 |
8.50 |
|
5.20 |
0.16 |
3.00 |
0.18 |
3.40 |
|
9.00 |
0.16 |
1.70 |
0.30 |
3.30 |
|
Correlation Data |
|
|
Parameter |
Data Observed |
|
N |
196 |
|
Range (U/L) |
1.6– 7.3 |
|
Regression |
Y = 0.981 + 0.08 |
|
Correlation |
r = 0.990 |
|
Sy.x |
0.15 |