Correlation A comparison of this method using an automated analyzer was carried out against a reference method based on the ferrosine reaction. The performance of Catachem Total Iron method was found essentially equivalent to the reference method used. Linear regression analysis produced the following information: |
|
Method Performance Characteristics Sensitivity: The sensitivity of this method is 0.00010-0.00012 absorbance units per µg/dL. Linear Range: In this method there is no significant nonlinearity over the range of 0-1200 µg/dL. Precision: Within-run and day-to-day precision is summarized below. |
|
*CV’s not meaningful when the average approaches zero. |
|
Intended Use For IN VITRO diagnostic use in the quantitative determination of Total Iron in serum using manual or automated applications.
Method Principle In acid buffer, transferrin bound iron dissociates into free ferrous (Fe++) and ferric (Fe+++) iron. Hydroxylamine hydrochloride present in the same acid reagent reduces the ferric iron to the ferrous state. The free ferrous ions react with ferene to form a complex with maximum absorbance at 600nm. The absorbance is directly proportional to the concentration of total iron in the serum sample. The reaction scheme illustrates the reaction that occurs in this method. |
|
TO CONTACT US:
CATACHEM, INC. 955 Connecticut Avenue Suite No. 4106 Bridgeport, CT 06607
Phone: 203-335-9277 Fax: 203-335-9977
Email: catachem@catacheminc.com
URL: www.catacheminc.com |
|
Total Iron |
Within-Run Precision |
Total Precision |
||
|
MEAN |
SD |
CV |
SD |
CV |
|
µg/dL |
µg/dL |
% |
µg/dL |
% |
|
25 |
2.60 |
* |
3.40 |
* |
|
477 |
7.40 |
1.50 |
8.20 |
1.70 |
|
948 |
5.10 |
0.50 |
14.80 |
1.50 |
|
Correlation Data |
|
|
Parameter |
Data Observed |
|
N |
93 |
|
Range (µg/dL) |
7-245 |
|
Regression |
Y = 0.989 + 2.0 |
|
Correlation |
r = 0.961 |
|
Sy.x |
9.9 |