Correlation

A comparison of Catachem Triglycerides method, using an automated analyzer and a reference method based on the GPO Trinder reaction resulted in the following regression statistics:

Preparation Of Working Reagent

Catachem Triglycerides Working Reagent is prepared by adding the required volume of distilled or deionized water to the required number of vials of Triglycerides reagent.

 

Reagent Storage And Stability

Catachem Triglycerides reagent is stored at 2-8°C. When stored as directed, this reagent is stable until the expiration date stated on the label. When prepared and stored as directed, the Working TG Reagent is stable for 3 days at room temperature (18-26°C) and for 30 days at  2-8°C.

 

Method Performance Characteristics

Sensitivity:  0.0008-0012 absorbance units per mg/dL.

Linear Range: 0-1000 mg/dL.

Precision: Within-run and day-to-day precision is summarized below:

Intended Use

For IN VITRO diagnostic use in the quantitative determination of Triglycerides in serum using manual or automated applications.

 

Method Principle

Serum Triglycerides are hydrolyzed by microbial Lipase to glycerol and free fatty acids. The resultant glycerol, in the presence of glycerol kinase (GK) ATP and Mg++ ions is phosphorilated to glycerol-1-phosphate. This latter metabolite is then oxidized by glycerophosphate oxidase (GPO) to produce hydrogen peroxide. The hydrogen peroxide thus produced is quantitatively determined by coupling 4-aminoantipyrine with                              N-ethyl-N-(2-hydroxy-3-sulphopropyl)-m-toluidine (TOOS) , where a quinonemine dye with maximum absorption at 550nm is produced. The following reaction scheme illustrates the reactions that occur in this method:

TO CONTACT US:

 

CATACHEM, INC.

353 Christian Street

Suite 2

Oxford, CT  06478

 

Phone:     203-262-0330

Fax:         203-262-9836

 

Email:

catachem@catacheminc.com

 

URL:

www.catacheminc.com

Triglycerides

Within-Run Precision

Total Precision

MEAN

SD

CV

SD

CV

mg/dL

mg/dL

%

mg/dL

%

69

1.10

1.60

1.50

2.10

386

1.12

1.30

2.90

0.70

714

3.50

0.50

7.40

1.00

Correlation Data

Parameter

Data Observed

N

118

Range (mg/dL)

29-715

Regression

Y = 1.036 + 5.8

Correlation

r = 0.998

Sy.x

7.00