Intended Use

For IN VITRO quantitative determination of Albumin in serum or plasma using manual or automated applications.

 

Method Principle

BCG dye at a pH of 4.2 and in the presence of a non-ionic surfactant Brij-35 specifically binds albumin to form a colored BCG-Albumin complex with maximum absorbance at 630nm. The increase in absorbance is directly proportional to the Albumin concentration present in the sample.

 

                                               

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Correlation

A comparison of this method using an automated analyzer and a reference method based upon the BCG reaction resulted in the following regression statistics.

Method Performance Characteristics

Sensitivity: The sensitivity of this method is 0.140-0.180 absorbance units per g/dL.

Linear Range: In this method there is no significant nonlinearity over the range of 0-6 g/dL.

Precision: Within-run and day-to-day precision is summarized below.

TO CONTACT US:

 

CATACHEM, INC.

353 Christian Street

Suite 2

Oxford, CT  06478

 

Phone:     203-262-0330

Fax:         203-262-9836

 

Email:

catachem@catacheminc.com

 

URL:

www.catacheminc.com

Albumin

Within-Run

Total

Mean

SD

CV

SD

CV

g/dL

g/dL

%

g/dL

%

1.2

0.04

3.50

0.05

4.30

3.4

0.05

1.40

0.07

2.00

5.6

0.06

1.10

0.06

1.00

Correlation Data

Parameter

Data Observed

N

197

Range

1.6-5.2 g/dL

Regression

Y= 1.039 - 0.12

Correlation

r = 0.999

Sy.x

0.09